Brightfield
Brightfield checklist#
- Make sure that the condenser is set to the correct position for the objective
- See the "Condenser for DIC" column
- Make sure everything (including your slide) along the lightpath is clean / free of dust
- If using DIC, use the air puffer to blow dust off the polarizer.
- Perform Kohler Alignment using the 10x objective
- See the Nikon guide for step-by-step instructions
- DIC only: Make sure the objective-specific DIC slider is inserted
- If using DIC, slide the polarizer in the lightpath. Otherwise keep it empty.
TwinCam DM severely reduces light throughput