Shut down#
Shutting down#
- Close NIS-Elements and wait about 10 seconds before turning off the scope body.
- Ensure the objectives are lowered all the way down
- Turn off scope body (before moving to next step, wait for light show/beeping to finish)
- Turn off X1 unit by rotating key counterclockwise
- Turn off controller for microscope
- Turn off laser box by rotating key counterclockwise
- Turn off surge protector (side switch)
- Cover the microscope
If you want to sync your data on the PSH network drive immediately, go to the computer Desktop and double-click on "BACKUP-X1". A black window will show up and automatically close when finished backing up the files.
Shutting down (live-cell)#
- Turn off the Tokai Hit unit by pushing the orange button.
- Close the Apollo valve for CO2 by rotating clockwise 90 degrees until the yellow handle is perpendicular to the tubing.
- Lower the objective as far as it'll go.
- Turn off the Bioptechs controller by pushing the On/Off button
- Remove the Bioptechs objective warmer and set back in box.
- Remove stage top incubator lid and set on air table.
- Remove water from reservoir using a syringe, and deposit into glass beaker labelled "waste".
- When no more water can be removed using the syringe, switch to using a KimWipe to soak up the remaining water.
- Inspect your sample adapter for oil (and clean if needed).
- See normal shut-down procedures